Chunk #21 — Methods — Measurements and data analysis — EEG data acquisition

Source
A stable pattern of EEG spectral coherence distinguishes children with autism from neuro-typical controls - a large case control study.
Embedded
yes

Text

Hz bandpass filtering and digitized at 256 Hz for subsequent analyses. All amplifiers were individually calibrated prior to each study. One other amplifier type was utilized for five patients with ASD (Bio-logic™, Bio-logic Technologies, Natus Medical Inc., 1501 Industrial Road, San Carlos, CA 04070 USA; 250 Hz sampling rate, 1 to 100 Hz bandpass) and one other amplifier type was utilized for 11 control subjects (Neuroscan™, Compumedics Neuroscan, 6605 West W.T. Harris Boulevard, Suite F, Charlotte, NC 28269 USA, 500 Hz sampling rate, 0.1 to 100 Hz bandpass). Data from these two amplifiers, sampled at other than 256 Hz. were interpolated to the rate of 256 Hz by the BESA 3.5™ software package. As the band-pass filter characteristics differed among the three EEG machines, frequency response sweeps were performed on all amplifier types so as to permit modification of data recorded with the Biologic and Neuroscan amplifiers to be equivalent to those gathered by the Grass amplifiers. This was accomplished by utilizing special software developed in-house by the first author using forward and reverse Fourier transforms [37].